Accu-Chek Guide error code E-7

Accu-Chek Guide

Error codes E-7

An electronic error occurred. 
Remove the batteries, press and hold either meter button for at least 2 seconds, then release the button and wait for at least 20 seconds before reinserting the batteries.
Perform a blood glucose or control test.